Immunoglobulin A, Individual Quant
Order Code: IGA
Order Form: A-1a General Lab or IPR Req
  Chemistry
6240 RCP
356-3527
Specimen
Plasma
Collection Medium:
Plasma separator tube
Alternate
Collection Media:
Call laboratory for additional acceptable specimen collection 
containers.
Minimum:
2 ml whole blood from light green top tube
Testing Schedule:
24 hrs/day, 7 days a week, including holidays.
Analytic Time:
2 hours (upon receipt in laboratory)
Reference Range:
70-400 mg/dl
Test
Limitations:
Criterion: Recovery within plus or minus 10% of initial value.

Icterus: No significant interference up to an I index of 60 
(approximate conjugated and unconjugated bilirubin concentration: 60 
mg/dl).

Hemolysis: No significant interference up to an H index of 1000 
(approximate hemoglobin concentration: 1000 mg/dl).

Lipemia (Intralipid): No significant interference up to an L index of 
3000). There is poor correlation between turbidity and triglycerides 
concentration.

There is no cross-reaction between IgA and IgG or IgM under the assay 
conditions.

Plasma with ill-defined characteristics should be subject to protein 
electrophoresis to identify a possible antigen excess (e.g. 
gammopathy).

A high-dose hook effect may occur at concentrations above 12,000 mg/dl 
(120 g/l).
Methodology:
Immunoturbidimetric
CPT Code:
82784
 
See also:
Cryoglobulin, Serum